Jörg Flock

X-Ray Fluorescence Spectroscopy for Laboratory Applications


Скачать книгу

excitation efficiency and the very small fluorescence yield for these elements (see Figure 2.4). On the other hand, due to the high fluorescence yield for heavier elements the ratio between fluorescence radiation and the spectral background is larger even if here the elastic scattering and consequently also the spectral background have a higher intensity. This is demonstrated in Figure 2.8 for the spectra already shown in Figure 2.7a. However, Figure 2.8 shows the direct measured, non-normalized spectra. Here it can be clearly seen that the background of the spectra of polymethylmethacrylate (PMMA) and Al are large compared to the fluorescence lines. On the other hand, the spectra of Ti and Ni have higher fluorescence intensities and therefore a lower spectral background.

Charts depicting the influence of the sample type on the intensity of the scattered radiation (y-axis in square root scale). Chart depicting the spectra of PMMA, Al, Ti, and Ni demonstrating the different scatter behavior (y-axis in square root scale).

      The intensity of the elastic scattered radiation depends on the direction of the scattered radiation. The scattering intensity of non-polarized radiation on a sphere with the diameter R and the scattering object in its center can be described by the following formula, Θ being the scatter angle:

      (2.7)equation

      As a result, the scatter intensity has a maximum in the forward and backward directions, while it is reduced perpendicular to the direction of irradiation.

      Inelastic scattering means a loss of energy of the scattered radiation compared to the original incident radiation. The energy is transferred to the scattering particle, most often an electron, which is then emitted as photoelectron. Because the conservation laws for both energy and momentum apply, the energy loss ΔE depends on the scatter angle Θ and on the energy E of the scattered radiation. For scattering on an electron this dependence is described by the following formula:

      (2.8)equation

      2.2.5 Detection of X-ray Spectra

      The radiation emitted by a sample is detected by an X-ray spectrometer. The spectrometer separates energetically different beam components and determines their intensity within narrow energy ranges; they can be, for example, individual element peaks as well as background areas with narrow energy ranges of a few electronvolts. In principle, two instrument types for spectrometry are in use where the dispersion takes place by different means:

       In the case of wavelength-dispersive spectrometers (WDSs – see Section 4.3.2), the separation of the radiation components takes place via a dispersive element, for example, a crystal, or, at low energies, by special multilayer structures, on which the fluorescence radiation of the sample is diffracted. The “reflection” of individual energies takes place only at defined diffraction angles, following Bragg's law, see Eq. (4.1).

       In the case of energy-dispersive spectrometers (EDSs – see Section 4.3.1.3), the dispersion is performed directly in the detector and its associated electronics. This generates an energy-dependent signal from each individual absorbed X-ray photon. By means of pulse-height analysis, the probability distribution of the photon energies absorbed in the detector can be generated, creating an image of the emitted spectrum.

      The intensity of this characteristic radiation depends on the number of atoms in the analyzed material, i.e. on their contents w. In a first approximation, therefore, the mass fraction w = ε·I, where ε is the element-dependent sensitivity, and I is the measured fluorescence intensity of the element under consideration. Unfortunately, the conditions are more complex than this, as all other elements in the sample influence through absorption and secondary excitation the measured fluorescence intensities. For a quantitative analysis this matrix influence has to be considered. It leads to complex corrections of the above linear dependency (see Section 5.5).

      The period from the beginnings of early X-ray spectrometry to today's powerful instrument technology has been long. The first stage was characterized by the development of sufficiently powerful instrument components and the development of basic mathematical models for matrix interaction. In the next step, the use of computing technology for data preparation as well as for instrument control was an important step for the establishment of the method for automated industrial use. Finally, the application areas of the method have been significantly expanded in the last 20 years by the availability of various X-ray optic elements as well as more powerful detectors.

      The development of X-ray fluorescence is briefly described by Niese (2007). The foundations for the use of X-ray spectrometry for element analysis were laid out by the discoveries of Moseley and Laue, and by using X-ray radiation for the screening of the human body for medical purposes. Experience was gained in making components such as X-ray tubes for the emission and X-ray films for the detection of the radiation. These were the preconditions for the initial use of X-ray spectrometry for element analysis.