exc, E em, T) is the emitted luminescence intensity.

      (2.56)equation

      The intensity absorbed by the sample is the difference between I 0 and I tr:

      If the absorption coefficient and the concentration of the absorbing defects are known, the absorption cross section σE exc can be obtained as:

      (2.58)equation

      Another spectroscopic parameter, frequently used to quantify the absorption probability, is the dimensionless oscillator strength f of an electric dipole transition of energy E between the initial, I, and final, II, electronic states:

      (2.59)equation

      (2.60)equation

      where n is the index of refraction of the medium.

      Due to the absorption process, (N 1 E exc) defects will be in the excited state during exposure of the sample to the excitation light, then a portion of them can decay radiatively, with a rate k r , thus originating the photon emission (luminescence) with energy E emE ex, while the remaining excited defects decay non‐radiatively, with a rate k nr that depends on temperature. The luminescence intensity is given by:

      where I lum(E em) is the emission lineshape determined by homogeneous and inhomogeneous contributions, in agreement with Eq. (2.55).

      The variation rate of the excited state population, N 1, depends on the absorption and decay processes, both radiative and non‐radiative, according to the following equation:

      1 The emission spectrum (PL spectrum) in which the intensity is measured as a function of E em for fixed E exc . This type of spectrum measures the shape and intensity of the band emitted by the sample.

      2 The excitation spectrum (PLE spectrum) measures the luminescence intensity, monitored at a fixed E em , as a function of E exc and represents the excitation efficiency of the luminescence spectrum.

      The time dependence of luminescence spectra is studied by using pulsed excitation. After the excitation of an ensemble of point defects with a light pulse, which produces a population of N 1(0) of the excited state, the light source is switched off (I 0 = 0) and N 1(t) decays according with:

      (2.64)equation

      where τ = 1/(k r + k nr ). Combining with Eq. (2.61), we obtain the expression of the luminescence time decay:

      The measure of the lifetime at low temperature, at which non‐radiative rates can be neglected, is only dependent on the radiative rate (τ r = 1/k r ); it allows to determine the oscillator strength by using the following equation, derived from the relation between Einstein's coefficient for spontaneous and stimulated emission:

      2.2.2 Site‐Selective Luminescence

      The use of tunable lasers in the time‐resolved luminescence setup is very advantageous to improve the sensitivity in the acquisition of emission spectra. On the other hand, it allows a selective excitation of defects, on which the site‐selective luminescence is based. To provide a theoretical background on this technique, we consider the absorption and luminescence lineshapes discussed in Section 2.1.5 accounting for the homogeneous and inhomogeneous parts. In fact, each defect contributes to the absorption through a homogeneous lineshape governed by the coupling between the electronic transition and both the band vibration and the localized modes. Under the hypothesis of linear coupling with a single localized phonon ℏΩ, the homogeneous lineshape is expressed by:

      (2.67)equation