Jörg Flock

X-Ray Fluorescence Spectroscopy for Laboratory Applications


Скачать книгу

10 wt% (cps) Steel 333 10.3 6 000 Aluminum 60 170 35 000 Polyethylene 15 1 300 170 000

Excited area (mm) Volume/analyzed mass for
Si in SiO2 Si in steel Fe in SiO2 Fe in steel Cd in aqueous solution
∅ 20 6 mm3 0.6 mm3 72 mm3 23 mm3 7500 mm3
16 mg 5 mg 190 mg 175 mg 7.5 g
∅ 0.3 0.001 mm3 0.0001 mm3 0.005 mm3 0.0015 mm3 1.7 mm3
0.003 mg 0.0016 mg 0.015 mg 0.04 mg 1.7 mg

       that in the case of surface contamination it is important to ensure that their thicknesses are small compared to the information depth and thus cannot influence the analytical result too much;

       that the surface roughness should be small against the information depth so that the absorption lengths of the fluorescence radiation are not influenced by topological effects – this influence can be reduced by rotating the sample during the measurement; or

       that for light matrices and high fluorescence energies, the information depth can exceed the sample thickness. Then the measured fluorescence intensity depends not only on the concentration of the analyte but also on the sample thickness. This is shown in Figure 3.3 for the intensities of Cd in polymer samples of different thicknesses. This problem can be solved by using the same amount of sample material for all analyzed samples, i.e. both measured and reference samples.

      

      3.2.3 Infinite Thickness

      The infinite thickness of a sample is another important parameter to consider, in particular for the analysis of layered samples (see Chapter 14). Like the information depth it depends on the element in question, other elements in the matrix, and the respective measuring geometry but it is largely independent of the spectrometer type. Typical dinfinite is assumed as three times dinformation, since the usable thickness range cannot be expanded to infinity, because the slope of the calibration curve for thick layers approaches asymptotically the infinite value. Infinite thickness does not mean that no other radiation can penetrate this layer – radiation of higher energy will not completely be absorbed in layers of this thickness. The infinite thickness, for example, of Si is about 15 μm, but only approximately 21% of the fluorescence radiation of Cu will be absorbed in this layer. In the matrix, backscattered Cu radiation can even enhance the fluorescence intensity of Si.

Chart depicting Cd intensities measured on polymer samples of different thicknesses, both measured and reference samples.

      Source: Courtesy of S. Hanning, FH Münster.

Analyzed volume limited by measurement geometry, because both the incident and the fluorescence radiation hit the sample at an angle less than 90 degrees.

      3.2.4 Contaminations

      For all preparation steps, the sample material comes into contact with other materials as well as with the laboratory environment. This can cause contaminations. These should, however, be avoided as far as possible in order not to influence the analytical result. Contaminations by the laboratory environment