usual, for spectroscopy, to use high‐pressure discharge lamps to enhance the continuous background and employ them in a large spectral range. In particular, lamps with Xe are used for UV‐Vis emission spectroscopy and with Deuterium for UV and up to vacuum‐UV absorption and emission.

       Laser [10]: various types of laser light are employed for spectroscopy due to their precise wavelength, spatial collimation, high intensity and time resolution in the pulsed regime [23]. These systems are realized with opportune cavities with reflecting walls to obtain the lasing effect.

      Gas lasers (e.g. He–Ne: 633 nm; Ar: 514.5 nm; Kr: 647 nm) employ an electric discharge to ionize a low‐pressure gas and induce the population inversion. Typically, they are characterized by continuous wave emission.

      Excimer lasers use mixture of gases [Xe, HCl, Ne (as buffer); Kr, F, He (as buffer)] at high pressure to obtain molecular complexes in the excited state (XeCl*, KrF*) through an electric discharge. During the excited state decay, a laser pulse is generated with duration in the window 5–20 ns. The population inversion is related to the disappearance of the ground state once the complex, which is unstable, dissociates.

      Dye lasers use organic molecules in solution as laser active medium with emission in the range 300–1000 nm depending on the dye. The emission range is tunable up to 30 nm and can be pulsed by pumping it with an excimer laser.

      Solid state lasers have been the first system used by Maiman in 1960 to prove coherent emission [24]. They include the first member of the family, the Ruby laser (Al2O3:Cr3+, 694.3 nm), the Nd‐YAG laser (1064 nm), and the Ti‐sapphire laser (Al2O3:Ti3+, ~800 nm, wide band). In these systems, a pumping lamp populates an excited state starting from which a long lifetime excited state is successively populated, inducing a population inversion and the following stimulated emission. The emission can be pulsed and a timescale of nanoseconds or down to femtoseconds can be reached. Furthermore, due to the high intensity of some of these systems, nonlinear effects can be obtained through nonlinear optics to generate harmonics or continuous very fast pulsed emission.

      Diode lasers are light emitting systems using semiconductors [23, 25]. They are solid structures of layered semiconductors with opportune bandgap tuning. This way, it is possible to obtain recombination of electrons and holes in a direct bandgap region. The charge injection by an external electric potential enables the emission of photons in this active region. By opportune combination of the constituting elements of the semiconductors (for example changing the x balance in Ga1−x Al x As), the emission can be engineered. The typical commercial diode lasers emit in the UV‐Vis‐IR range.

      1.3.3 Dispersion Elements: Gratings and Resolution Power

      (1.114)equation

      where m is an integer, giving the order of interference [1]. The presence of more slits implies that between two adjacent maxima, given by different values of m, there are minima of intensity for

      (1.115)equation

      These equations can be used to show that the angular distance between the maxima of interference of two wavelengths at distance Δλ is given by

      (1.116)equation

      giving the angular dispersion of a grating. This depends on the distance between the slits, usually reported as (number of lines)/mm, and it increases on increasing the lines/mm constituting the grating (lower value of d) and on increasing the order m of interference. If a monochromator of focal length f is considered [2, 22], the linear dispersion is obtained

      (1.117)equation

      yielding the spatial distance Δx, known as linear distance, on a planar screen of the interference maxima of two wavelengths separated by Δλ, and showing that this distance increases on increasing the focal length of the monochromator. Finally, in order to be able to separate the interference maxima of two adjacent wavelengths, the condition usually applied is that the minimum intensity of one wavelength coincides with the maximum intensity of the other. This condition gives the resolving power of a grating:

      (1.118)equation

      this is the minimum difference of wavelengths that a grating with N lines can resolve at the interference order m. This feature depends on the number N of slits illuminated and does not depend on their relative distance d. All of the reported features are fundamental for the good choice of a grating to be used in a monochromator.